If exploited, the vulnerability could allow device settings to be modified and/or interrupt the instrument startup process. The ability to exploit this vulnerability is limited, as it would require a direct physical intrusion into the instrument and would require the user to gain access to and tamper with an internal port or memory.
Abbott has reviewed the level of potential impact of this vulnerability on product performance and safety, and has taken appropriate steps to mitigate this issue in its latest software update, v7.1.
CISA Number: N/A
CVE Number: CVE-2023-47262
Vendor Website: Vulnerability Information
Additional Information: See vendor website linked above for additional information